0-10V Analogue interface (standard)

Standard 0-10V remote analogue interface, with potential free contacts for the instructions and statuses

As a standard, our power supplies can be remotely controlled through a 0-10V analogue interface. The interface consists in a single DB-25 connector, located on the rear panel of the power supply.

The remote control cable must be shielded, and we recommend to install a length shorter than 3m.

Detailed specification

Pin
Signal
Signal description
I/O
Impedance
1
HV-Off control
Produced by a fleeting opening from the digital ground reference (pin16)
input
20 kΩ
2
Fault status
Internal Fault: 0V = An internal fault has been detected; +15V = No fault
output
100 Ω
3
Interlock status
0V = External interlock is open; +15V = External interlock is closed
output
475 Ω
4
HV-On control
Produced by a fleeting closing to the digital ground reference (pin16)
input
20 kΩ
5
Output voltage measurement
0V to 10V = 0% to 100% of output voltage
output
475 Ω
6
Output current measurement
0V to 10V = 0% to 100% of output current
output
475 Ω
7
Inhibit control
Activated by a digital signal between +5V to +24V
input
8
Remote control
Open contact = Local control mode; Closed contact= Remote control mode
input
20 kΩ
9
Not Connected
10
Mains Monitor
Generates a fault if the mains input is failing: 0 = Mains is fine ; +15V: Mains is failing
output
100 Ω
11
Output power measurement (option)
0V to 10V = 0% to 100% of output power (option)
output
475 Ω
12
Local output voltage setting
Copy of the output voltage setting on the front panel potentiometer. 0V to 10V corresponds to 0% to 100% of the setting.
output
1 MΩ
13
Local output current setting
Copy of the output current setting on the front panel potentiometer. 0V to 10V corresponds to 0% to 100% of the setting.
output
1 MΩ
14
Remote output current setting
0V to 10V = 0% to 100% of output current
input
15 kΩ
15
+10V Reference
+10V reference for anologue signals, max current : 5mA
output
2.7 Ω or 15 kΩ
16
0V reference (digital signals)
0V ground reference for digital signals
output
17
Remote output voltage setting
0V to 10V = 0% to 100% of output voltage
input
15 kΩ
18
Regulation mode status or End of Charge status
For a DC power supply: Open contact = Current Regulation; +24V = Voltage Regulation For a Capacitor charging power supply: Open contact = Capacitor is being charged; +24V = End of charge
output
100 Ω
19
HV-On status
0V = HV output is disabled (HV Off); +15V = HV output is enabled (HV On)
output
100 Ω
20
0V reference (analogue signals)
0V ground reference for analogue signals
output
21
Not Connected
22
Not Connected
23
Not Connected
24
External Interlock
Connect this pin to the digital to the 0V reference for digital signals (pin 16) to close the interlock.
input
500 Ω
25
+10V Reference or Remote output power setting (option)
Standard : +10V reference for anologue signals, max current : 5mA Option: 0V to 10V = 0% to 100% of output power